KAIWEETS KE2500

KAIWEETS KE2500 Professional Insulation Resistance Tester User Manual

Modail: KE2500

1. Fiosrachadh Sàbhailteachd

Before operating the KAIWEETS KE2500 Insulation Resistance Tester, please read and understand all safety warnings and operating instructions. Failure to follow these instructions may result in electric shock, fire, or serious injury.

  • Cum an-còmhnaidh ri còdan sàbhailteachd ionadail is nàiseanta.
  • This device is compliant with IEC/EN 61010-1 CAT III 1000V / CAT IV 600V safety standards.
  • Do not use the tester if it appears damaged or if the test leads are compromised.
  • Ensure the device is turned off and disconnected from any circuit before changing test leads or batteries.
  • Wear appropriate personal protective equipment (PPE), including insulated gloves and safety glasses, when performing electrical tests.
  • Avoid contact with live circuits. High voltages are present during insulation resistance testing.
  • Do not operate the tester in wet conditions or in the presence of explosive gases or dust.
  • Keep hands and fingers behind the probe barriers during testing.

2. Toradh thallview

The KAIWEETS KE2500 is a professional digital insulation resistance tester designed for precise measurements in various electrical maintenance and safety verification tasks. It offers a wide range of test voltages and advanced diagnostic capabilities.

Prìomh fheartan:

  • 5-Range Insulation Testing: 100V, 250V, 500V, 1000V, 2500V DC.
  • Wide Resistance Range: 0.1MΩ to 200GΩ.
  • Diagnosachd adhartach: One-touch Absorption Ratio (DAR) and Polarization Index (PI) calculation.
  • Intelligent PASS/FAIL Evaluation: Customizable thresholds for quick quality control.
  • Ioma-ghnìomhachd: Tomhais AC/DC Voltage (up to 750V AC / 1000V DC) and Continuity (buzzer for <50Ω).
  • Dealbhadh càirdeil don neach-cleachdaidh: Backlit LCD, Data Hold, 15-minute Auto Shutdown, Max/Min recording.

Dè tha air a ghabhail a-steach:

  • KAIWEETS KE2500 Insulation Tester (x1)
  • 1.5V AA Batteries (x8)
  • Test Leads (x2)
  • Crocodile Alligator Clips (x2)
  • Poca giùlain (x1)
  • Leabhar-làimhe Cleachdaiche (x1)
KAIWEETS KE2500 Insulation Resistance Tester with accessories

Image: KAIWEETS KE2500 Insulation Resistance Tester and included accessories.

Bhidio: Thar-shealladhview of the KAIWEETS Professional Insulation Resistance Tester KE2500, demonstrating its features and use.

3. Suidhich

Stàladh bataraidh:

  1. Dèan cinnteach gu bheil an deuchainniche dheth.
  2. Lorg an roinn bataraidh air cùl an inneil.
  3. Fosgail an còmhdach roinn bataraidh.
  4. Cuir a-steach 8 bataraidhean AA, a’ cumail sùil air a’ phòlaireachd cheart (+/-) mar a tha air a chomharrachadh taobh a-staigh an roinn.
  5. Dùin an còmhdach roinn bataraidh gu tèarainte.
KAIWEETS KE2500 with battery compartment open

Image: The KAIWEETS KE2500 with its battery compartment open, showing the placement for 8 AA batteries.

A’ ceangal uèirichean deuchainn:

  1. Connect the red test lead to the 'LINE' terminal and the black test lead to the 'EARTH' terminal for insulation resistance tests.
  2. Airson voltage measurements, connect the red test lead to the 'V' terminal and the black test lead to the 'G' (Ground) terminal.
  3. For continuity tests, connect the red test lead to the 'CONTINUITY' terminal and the black test lead to the 'G' (Ground) terminal.
  4. Attach the alligator clips to the test leads as needed for secure connections.
KAIWEETS KE2500 showing test lead connections

Image: Close-up of the KAIWEETS KE2500 showing the input terminals for test leads, labeled LINE, V, G, EARTH, and CONTINUITY.

4. Stiùireadh Obrachaidh

Cumhachd air / dheth:

Rotate the central dial from 'OFF' to any desired measurement function to power on the device. Rotate back to 'OFF' to power off. The device also features a 15-minute auto-shutdown to conserve battery life.

Deuchainn an aghaidh insulation:

  1. Ensure the test leads are correctly connected to the 'LINE' and 'EARTH' terminals.
  2. Rotate the dial to select the desired test voltage (100V, 250V, 500V, 1000V, no 2500V).
  3. Ceangail na uèirichean deuchainn ris a’ chuairt no ris a’ phàirt fo dheuchainn.
  4. Press the 'TEST' button to initiate the insulation resistance measurement. The display will show the resistance value.
  5. Release the 'TEST' button to stop the measurement.
KAIWEETS KE2500 performing insulation resistance test

Image: The KE2500 displaying an insulation resistance measurement, showing the range up to 200GΩ.

Timed Measurement Function:

  1. After selecting the test voltage, press the 'TIMER' button.
  2. Use the 'HOLD' and 'MAX/MIN' buttons to adjust the desired test duration (e.g., 1 minute, 5 minutes).
  3. Press 'TEST' to start the timed measurement. The device will automatically stop after the set time.
KAIWEETS KE2500 showing timed measurement function

Image: The KE2500 displaying a timed insulation resistance test, showing the elapsed time.

DAR/PI Calculation:

The Absorption Ratio (DAR) and Polarization Index (PI) are indicators of insulation quality. The KE2500 can calculate these automatically.

  1. Perform an insulation resistance test as described above.
  2. Press the 'DAR/PI' button. The device will display the calculated DAR or PI value.
  3. DAR is calculated as the ratio of 60-second insulation resistance to 15-second insulation resistance.
  4. PI is calculated as the ratio of 10-minute insulation resistance to 1-minute insulation resistance.
KAIWEETS KE2500 displaying DAR/PI calculation

Image: The KE2500 showing the DAR/PI calculation function, indicating the ratio of resistance measurements over time.

Threshold Comparison (PASS/FAIL):

  1. Press the 'COMP' button to enter comparison mode.
  2. Use the 'HOLD' and 'MAX/MIN' buttons to set your desired insulation resistance threshold.
  3. Perform an insulation resistance test. The display will indicate 'PASS' or 'FAIL' based on your set threshold.
KAIWEETS KE2500 showing PASS/FAIL threshold comparison

Image: The KE2500 displaying a PASS/FAIL result based on a user-defined threshold for insulation resistance.

AC / DC Voltage Tomhais:

  1. Connect the test leads to the 'V' and 'G' terminals.
  2. Rotate the dial to the 'AC/DC-V' position.
  3. Connect the test probes to the circuit to measure AC or DC voltage. The device automatically detects AC or DC.
KAIWEETS KE2500 measuring AC voltage

Image: The KE2500 measuring AC voltage in an electrical panel, showing the display with AC voltage leughadh.

KAIWEETS KE2500 measuring DC voltage

Image: The KE2500 measuring DC voltage on a car battery, showing the display with DC voltage leughadh.

Deuchainn leantainneachd:

  1. Connect the test leads to the 'CONTINUITY' and 'G' terminals.
  2. Rotate the dial to the 'Ω' position.
  3. Connect the test probes to the circuit or component. A buzzer will sound if the resistance is less than 50Ω.
KAIWEETS KE2500 performing continuity test

Image: The KE2500 performing a continuity test, showing the display with a low resistance reading and indicating continuity.

Data Hold, Max/Min, and Backlight:

  • Press 'HOLD' to freeze the current reading on the display. Press again to release.
  • Press 'MAX/MIN' to cycle through maximum, minimum, and average readings during a measurement session.
  • Press the backlight button (often combined with 'HOLD' or a separate button) to illuminate the LCD for low-light conditions.
KAIWEETS KE2500 with backlit LCD display

Image: The KE2500 with its backlit LCD display illuminated, showing clear visibility of readings.

5. Cumail suas

Glanadh:

Wipe the device with a dry, soft cloth. Do not use abrasive cleaners or solvents. Ensure the device is off and disconnected from any power source before cleaning.

Ath-nuadhachadh bataraidh:

When the low battery indicator appears on the display, replace all 8 AA batteries promptly to ensure accurate readings and proper operation. Refer to the 'Battery Installation' section for detailed steps.

Stòradh:

Store the tester in its carrying bag in a cool, dry place away from direct sunlight and extreme temperatures. If storing for extended periods, remove the batteries to prevent leakage.

6. A ’dèanamh trioblaid

  • Gun Taisbeanadh/Cumhachd Air: Thoir sùil air stàladh a’ bhataraidh agus dèan cinnteach nach eil na bataraidhean air an leigeil ma sgaoil. Cuir an àite iad ma tha sin riatanach.
  • Leughaidhean neo-mhearachdach: Ensure test leads are securely connected and not damaged. Verify the correct measurement function and voltage range are selected. Clean the test probes if they are dirty.
  • Taisbeanadh 'OL' (Cus Luchd): This indicates the measured value is outside the device's range. For insulation resistance, it means the resistance is too high to measure. For voltage, it means the voltage exceeds the maximum input.
  • Gun Bhùidsear Leantainneachd: Ensure the device is in continuity mode and the resistance is below 50Ω. Check test leads for proper connection.

7. Sònrachaidhean

SònrachadhLuach
BrandKAIWEETS
Modail ÀireamhKE2500
Stòr cumhachd8 Bataraidhean AA (air an toirt a-steach)
Raon Resistance Insulation0.1MΩ to 200GΩ
DC Test Voltages100V, 250V, 500V, 1000V, 2500V
AC Voltage TomhaisSuas gu 750V AC
DC Voltage TomhaisSuas gu 1000V DC
Crannchur Leantainneachd<50 gu
Rangachadh SàbhailteachdIEC/EN 61010-1 CAT III 1000V / CAT IV 600V
Auto Shutdown15 mionaid
Cuideam nì3.53 not (1600 gram)
Meudan toraidh11.81 x 9.06 x 2.76 òirleach

8. Barantas agus Taic

KAIWEETS provides a 36-month after-sales service for the KE2500 Insulation Resistance Tester. For any technical assistance or inquiries, please contact KAIWEETS customer support. Round-the-clock lifetime technical support is available to ensure your satisfaction and proper use of the product.

For further information or support, please visit the official KAIWEETS weblàrach-lìn no cuir fios chun t-seirbheis teachdaiche aca gu dìreach.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.