AmScope T490
AmScope T490 Compound Trinocular Microscope
Leabhar-làimhe Stiùireadh Cleachdaiche
Ro-ràdh
This manual provides detailed instructions for the setup, operation, and maintenance of your AmScope T490 Compound Trinocular Microscope. Please read this manual thoroughly before using the instrument to ensure proper function and longevity.
Fiosrachadh Sàbhailteachd
- Làimhsich am miocroscop gu faiceallach an-còmhnaidh. Seachain buaidhean no crithidhean obann.
- Do not disassemble the microscope. Refer servicing to qualified personnel.
- Ensure the power cord is securely connected and the voltage matches your local supply.
- Unplug the microscope before cleaning or performing any maintenance.
- Avoid direct exposure of the eyes to the light source.
- Keep the microscope away from direct sunlight, high temperatures, dust, and corrosive chemicals.
Dè th' anns a' Bhocsa?
Dearbhaich gu bheil na pàirtean uile a tha air an liostadh gu h-ìosal an làthair sa phacaid agad:
- AmScope T490 microscope with double-layer mechanical stage
- WF10x eyepieces, 23mm, one pair
- 4x DIN achromatic objective, 20mm
- 10x DIN achromatic objective, 20mm
- 40xS DIN achromatic objective, 20mm
- 100xS (oil) DIN achromatic objective, 20mm
- Immersion oil, one bottle
- (3) Color filters; blue, green, and yellow
- Spare halogen bulb
- fuse a bharrachd
- Còmhdach dust
- Corda cumhachd
- Stiùiridhean (an leabhar-làimhe seo)
Suidhich
Follow these steps to set up your AmScope T490 microscope:
- A' dì-phapadh: Thoir a h-uile pàirt a-mach às a’ phacaid gu faiceallach. Cùm am pasgan tùsail airson còmhdhail no stòradh san àm ri teachd.
- Suidheachadh: Place the microscope on a stable, level surface, away from direct sunlight and heat sources.
- Stàladh Ceann: Loosen the head locking screw on the microscope body. Gently slide the trinocular head onto the top of the microscope body. Ensure it is fully seated and then tighten the locking screw.
- Stàladh sùil-phìos: Insert the WF10x eyepieces into the eyepiece tubes of the trinocular head.
- Stàladh Amasan: Rotate the revolving nosepiece until the lowest power objective (4x) is positioned over the stage. Carefully screw each objective into the nosepiece in increasing order of magnification (4x, 10x, 40xS, 100xS).
- Ceangal cumhachd: Connect the power cord to the microscope's power input and then to a suitable electrical outlet (110V).

Figure 1: AmScope T490 Microscope with included accessories, showing the main unit, eyepieces, objectives, and various filters.
Stiùireadh Obrachaidh
1. Powering On and Illumination Adjustment
- Turn on the power switch located on the base of the microscope.
- Adjust the light intensity using the rheostat knob.
- The 1.25 NA Abbe condenser can be adjusted using its rack-and-pinion control system to optimize the distance of light from the stage.
- The iris diaphragm on the condenser controls the amount of light illuminating the specimen. Adjust it for optimal contrast and resolution.
2. A’ cur sampall
- Place the specimen slide on the double-layer mechanical stage.
- Secure the slide using the stage criomagan.
- Use the X-Y coaxial controls to move the slide and position the desired area of the specimen directly under the objective lens. The stage has 1mm divisions for precise positioning.

Figear 2: Taobh view of the AmScope T490, highlighting the mechanical stage and focus knobs.
3. Fòcas
- Begin with the lowest power objective (4x) rotated into position.
- Cleachd an cnap fòcas garbh to bring the specimen into approximate focus. Rotate the knob to move the stage suas no sìos.
- Cho luath ‘s a bhios tu ag amas gu garbh, cleachd an cnap fòcas grinn for precise focusing. The fine focus divisions are 0.002mm.
- The focus knob tension control prevents the stage from drifting out of focus.
- A stage stop is present to prevent the stage from contacting the objectives and damaging slides.
4. Atharrachadh Mheudachaidh
- Once focused at a lower power, rotate the revolving nosepiece to select a higher power objective (10x, 40xS, 100xS).
- The 40xS (spring) and 100xS (spring, oil) objectives are spring-loaded to prevent damage to the slide or objective.
- Only minor adjustments with the fine focus knob should be needed when switching between objectives due to the parfocal design.
5. Adjusting Eyepieces (Siedentopf Head)
- Astar eadar-mholaidh: Adjust the sliding head to match the distance between your eyes (interpupillary range of 55 to 75mm). Look through both eyepieces and adjust until you see a single, clear circular field of view.
- Atharrachadh Diopter: If one eye is stronger than the other, use the diopter adjustment ring on one of the eyepiece tubes to compensate. Focus with your dominant eye first, then adjust the diopter for the other eye without re-focusing the main coarse/fine knobs.
- The head has a fixed 30-degree vertical inclination for comfortable viewing and can rotate 360 degrees for shared viewing.

Figear 3: Dùin suas view of the trinocular head with WF10x eyepieces and the vertical camera mount.
6. Using the 100x Oil Immersion Objective
- When using the 100x objective, a drop of immersion oil must be placed on the specimen slide directly over the area to be viewdeas.
- Carefully lower the 100x objective into the oil drop until it makes contact.
- Use only the fine focus knob for focusing with the 100x objective.
- After use, always clean the 100x objective and the slide immediately with lens paper and a suitable cleaning solution to remove all oil residue.
7. Using the Trinocular Port
- The vertical trinocular port accepts a camera with a 23mm or C-Mount adapter (camera sold separately).
- Refer to your camera's manual for specific installation and operation instructions.

Figear 4: Aghaidh view of the AmScope T490, showing the objective turret and stage from a direct perspective.
Cumail suas
1. Glanadh
- Co-phàirtean Optaigeach: Use only specialized lens cleaning solutions and lens paper for cleaning objectives and eyepieces. Gently wipe in a circular motion from the center outwards. Never use abrasive materials.
- Corp: Clean the microscope body with a soft, damp cloth. Avoid using organic solvents.
- Always clean the 100x oil immersion objective immediately after use to prevent oil from drying on the lens.
2. Stòradh
- Nuair nach eilear ga chleachdadh, còmhdaich am miocroscop leis a’ chòmhdach duslach a tha air a thoirt seachad gus a dhìon bho dhuslach is sprùilleach.
- Stòraich am miocroscop ann an àite fionnar, tioram, air falbh bho sholas dìreach na grèine agus teòthachd anabarrach.
3. Bulb and Fuse Replacement
- Always unplug the microscope before attempting to replace the halogen bulb or fuse.
- Refer to the diagrams in the original product packaging for specific locations of the bulb and fuse compartments.
- Use only the specified replacement parts (6V/20W halogen bulb, spare fuse).
Fuasgladh thrioblaidean
| Trioblaid | Adhbhar a dh’fhaodadh a bhith ann | Fuasgladh |
|---|---|---|
| Gun soillseachadh | Power cord not connected; Power switch off; Bulb burnt out; Fuse blown. | Check power connection; Turn on power switch; Replace bulb; Replace fuse. |
| Ìomhaigh neo-shoilleir/doilleir | Improper focusing; Objective lens dirty; Eyepieces dirty; Condenser/diaphragm not adjusted correctly; Immersion oil not used with 100x objective. | Adjust coarse/fine focus; Clean lenses; Adjust condenser and iris diaphragm; Apply immersion oil for 100x objective. |
| Spotan dorcha ann an raon view | Dust on eyepieces, objectives, or slide. | Clean eyepieces, objectives, and slide. |
| Stage drifts out of focus | Focus knob tension too loose. | Adjust focus knob tension control. |
Sònrachaidhean
| Ceann | Compound Trinocular, Siedentopf |
| Meudachadh | 40x-1000x |
| Port Trinocular | 23mm or C-Mount |
| Sùilean | WF10x |
| Amasan | 4x, 10x, 40xS, 100xS (oil) DIN Achromatic |
| Stage | Double-layer Mechanical, 5-1/2 x 5-1/2 inches (140 x 140mm) |
| Suidheachadh solais | Transmitted (lower) |
| Condenser | 1.25 NA Abbe |
| Diaphragm | Iris |
| Stòr Solais | Halogen with rheostat, 6V/20W |
| Seòrsa Soillseachadh | Brightfield |
| Cumhachd | 110V |
| Interpupillary Range | 55 gu 75mm |
| Raon fòcas | 1-3/16" (30mm) |
| Fine Focus Divisions | 0.002mm (0.0000787") |
| Meudan toraidh | 17.45 x 9.77 x 20.33 òirleach |
| Cuideam nì | 17 not |
Barantas agus Taic
For warranty information and technical support, please refer to the documentation provided with your purchase or visit the official AmScope website. United Scope manufactures microscopy equipment and accessories under the brand name AmScope. The company, founded in 1996, is headquartered in Irvine, CA.